應用案例Case
|
|
Laboratory Autoclave
SN200C/210C/300C/310C/500C/510C,SQ500C/510C
The highest temperature of 135°C
Upwards opening lid
|
|
|
|
|
|
|
Product code
|
241005/241054
|
241006/241055
|
241007/241056
|
241008/241057
|
Model*1
|
SN200C/SN210C
|
SN300C/SN310C
|
SN500C/SN510C
|
SQ500C/SQ510C
|
System
|
Automatic high-pressure steam sterilization system for laboratory use ※2
|
Perfor-
mance
|
Range of working temperatures
|
45-135°C: 45-80°C for preheating, 45-60°C for keep worm, 65-100°C for melting, 105-135°C for sterilization
|
Max. working pressure
|
0.255MPa
|
Ambient temperature
|
5 - 35°C
|
Lid
|
Manual upward opening with an interlock for safety
|
Main unit
|
Heater
|
100V, 1.2kW (600W 2 units)
|
100V, 1.6kW (800W 2 units)
|
100V, 1.9kW (950W 2 units)
|
100V, 2kW (1000W 2 units)
|
Exhaust valve
|
An exhaust valve and a slow release valve
|
Connection ports for optional accessories
|
1/4" female thread for, sample temp. sensor, chamber temp. sensor, and pressure sensor (branching from the solenoid valve tubing) Total: 3 ports
|
Cooling fan
|
Axial fan motor
|
Controller
|
Temperature setting, control and display
|
PID control by microprocessor, digital setting (by UP/DOWN keys) and display
|
Timer and its resolution
|
0 or 1 minute to 99 hours 59 minutes in increments/decrements of 1 minute
|
Operation mode
|
Instrument sterilization course, liquid sterilization course, sterilization and keep/warm course, melting and keep/warm course, customer-programmed course
|
Other features
|
Key locking, presetting, saving, preheating, forced cooling, sample temperature sensor (option), memory locking, saving the history of up to 20 errors, indication of accumulated working hours, indication of present time, on-off beeping setting
|
Safety measures
|
Sensor malfunctioning, SSR short-circuit, broken heater wire, prevention of idle heating (liquid expansion type), alarm against the absence of a drain bottle, failure in locking the cover, memory malfunctioning, pressure relief valve (0.255 MPa)
|
Classification of pressure vessel
|
Small pressure vessel (installation report to authorities is not required)
|
Specifications
|
External dimensions(mm) ※3
|
Small pressure vessel (installation report to authorities is not required)
|
Effective internal dimensions (mm)
|
400(W)× 530(D)× 848(H)
|
460(W)× 590(D)× 848(H)
|
460(W)× 590(D)× 1,058(H)
|
520(W)× 660(D)× 876(H)
|
Effective Capacity
|
20L
|
32L
|
47L
|
50L
|
Weight
|
Approx. 65 kg
|
Approx. 75 kg
|
Approx. 85 kg
|
Approx. 95 kg
|
Power supply
|
SN200:
100V-120V AC 12.5A-15A
|
SN300:
100V-120V AC 16.5A-20A
|
SN500:
100V-120V AC 19.5A-23.5A
|
SQ500:
100V-120V AC 20.5A-24.5A
|
SN210:
200V-240V AC 6.5A-7.5A
|
SN310:
200V-240V AC 8.5A-10A
|
SN510:
200V-240V AC 10A-12A
|
SQ510:
200V-240V AC 10.5A-12.5A
|
Accessories
|
Stainless Basket (mesh type)
OSM-60(205mm diameter × 204mm deep) × 2 for SN200/210 OSM-70(262mm diameter × 204mm deep) × 2 for SN300/310
OSM-70(262mm diameter × 204mm deep) × 3 for SN500/510 OSQ-90(332mm diameter × 195.5mm deep) × 2 for SQ500/510
|
Steam cup × 1, drain bottle × 1, Drain board × 1, Chemical indicator × 1 set (30 pieces)
|
Power plug is not provided.
*1 Abbreviation (J) is Operation panel board in Japanese and (E) is in English.
*2 The autoclaves are not for medical use.
*3 External dimensions exclude protrusions.
|
Optional accessories
|
|
In addition to absorbent gauzes, laboratory garments and culturing instruments, a wide selection of basket and containers for laboratory waste are available.
|
|
Stainless bucket
|
Product code
|
Description
|
Model
|
Corresponding models
|
|
241087
|
Basket (mesh type)
|
OSM-60
|
For SM/SN/SP200C
|
|
241088
|
OSM-70
|
For SM/SN300C/500C/SP300C
|
*1
|
241089
|
OSM-80
|
For SM/SN/SP500C
|
*1
|
241099
|
OSQ-90
|
For SQ500C
|
|
241093
|
Basket (mesh type) with stacking fitting
|
OSQ-30
|
For SM/SN/SP200C, with two fittings/set
|
*1
|
241092
|
OSQ-40
|
For SM/SN/SP300C, with two fittings/set
|
*1
|
241091
|
OSQ-50
|
For SM/SN/SP500C, with three fittings/set
|
*1
|
241090
|
OSR-40
|
For SQ500C, with two fittings/set
|
|
241096
|
Basket (mesh type) with a mesh shelf
|
OSQ-60
|
For SM/SN/SP200C, with 1 mesh shelf
|
*1
|
241095
|
OSQ-70
|
For SM/SN/SP300C, with 1 mesh shelf
|
*1
|
241094
|
OSQ-80
|
For SM/SN/SP500C, with 2 mesh shelves
|
*1
|
241097
|
OSR-50
|
For SQ500C, with 1 mesh shelf
|
|
241083
|
Stainless solid basket
|
OSR-10
|
For SM/SN/SP200C
|
*1
|
241084
|
OSR-20
|
For SM/SN/SP300C/500C
|
*1
|
241098
|
OSR-60
|
For SQ500C
|
|
241150
|
Stainless bucket
|
OSN10
|
For SM/SN/SP200C
|
|
241151
|
OSN12
|
For SM/SN/SP300C/500C
|
|
241152
|
OSN14
|
For SQ500C
|
|
241153
|
Chamber temperature sensor
|
OSR-70
|
For SN/SQ, Type T thermocouple, 3 pieces/set
|
|
241154
|
Sample temperature sensor
|
OSN16
|
For SN/SQ, Type T thermocouple, 1 piece
|
|
241155
|
Pressure gauge
|
OSN18
|
For SN/SQ, external installation
|
|
241156
|
Recorder
|
RM10C
|
6-point type, temp. and pressure recording external installation
|
|
241157
|
External alarm output terminal
|
OSN20
|
For SN/SQ
|
|
241158
|
Temperature output terminal
|
OSN22
|
For SN/SQ, 4 - 20 mA
|
|
241159
|
Time-up output terminal
|
OSN24
|
For SN/SQ, relay, contact output
|
|
*1: Can be also used with conventional models SM200C/500C/510C and SP200C/300C/500C/510C.
|
|
|